Automobile interior inspection scheme

The headspace method uses a headspace sampler and gas chromatograph. The sampling method is relatively simple, and the sampling is completed in the headspace sampler. 2 grams of samples were taken from the test sample and put into a 20ml headspace bottle, which was kept warm and balanced at 120℃ for 60min. Then a certain amount of gas was extracted from the headspace bottle with a syringe, and analyzed by gas chromatograph and FID detector.

1.Test basis

This scheme is based on THE European standard VDA277 in ISO/TS16949, and is also a common method for measuring benzene hydrocarbon volatile organic compounds in European automotive enterprises.

2.Detection indicators

Nonmetallic materials in automotive interior include benzene, toluene, xylene, ethylbenzene, styrene, formaldehyde, tetradecane, etc.

3.Instrument preparation

4.Experimental conditions

Inlet temperature: 200℃
Column temperature: initial temperature 50℃, initial temperature 5min, heating rate 12℃/min, final temperature 200℃, final temperature 3min
Detector temperature: 250℃
Carrier gas: helium, flow rate is about 30 mL /min
Combustion gas: hydrogen, flow rate of about 30 mL /min
Auxiliary gas: air, flow rate is about 300ml/min

5.Atlas data


We are a branch of Kejie holding group, in charge of the overseas marketing of Kejie products, providing to customers with chromatograph, mass spectrometry, spectrum and other analytical scientifc instruments and professional services all around the world.Welcome to visit our website: www.changior.com or contact us via email: sales@changior.com